U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) CURRENT ACCEL VR, 36J; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) CURRENT ACCEL VR, 36J; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD1215-36
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/01/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event or Problem Description
It was reported that the patient presented via remote transmission with the implantable cardioverter defibrillator at elective replacement indicator.Based on longevity estimates, premature battery depletion was suspected.The physician explanted and replaced the device with no complications.Further investigation revealed, that the programmer had overestimated the battery longevity and that the device battery depletion was normal.The patient was stable and doing well.
 
Additional Manufacturer Narrative
The implantable cardioverter defibrillator was received at elective replacement indicator due to a reported event of diagnostic anomaly.The longevity calculation showed that the battery depletion was normal, but the estimate provided by the programmer was incorrect.Bench testing on the device showed values to be normal with no anomalies aside from procedural damage.The reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CURRENT ACCEL VR, 36J
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key8426485
Report Number2017865-2019-03412
Device Sequence Number2051925
Product Code NVZ
UDI-Device Identifier05414734502634
UDI-Public05414734502634
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2011
Device Model NumberCD1215-36
Device Lot Number3140684
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/01/2019
Supplement Date Received by Manufacturer07/29/2019
Initial Report FDA Received Date03/16/2019
Supplement Report FDA Received Date07/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
-
-