| Catalog Number |
122128048 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Joint Dislocation (2374); No Code Available (3191)
|
| Date of Event |
02/25/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.(b)(4).
|
| |
|
Event or Problem Description
|
|
Patient was revised to address dislocation.Constrained liner.Doi: (b)(6) 2012.Dor: (b)(6) 2019; left hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary = > no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|