• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MELSUNGEN AG INTROCAN SAFETY®; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MELSUNGEN AG INTROCAN SAFETY®; INTRODUCER, CATHETER Back to Search Results
Model Number 4251652-02
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Exemption number e2016018 b.Braun inc.(bbmi) (importer) is submitting the report on behalf of b.Braun melsungen ag (manufacturer).This report has been identified as event four of b.Braun medical inc.Internal report number (b)(4) the investigation into this reported event is ongoing.A follow-up report will be submitted when the results of the investigation are available.
 
Event or Problem Description
As reported by user facility: event 4: catheters have a visible plastic covering the needle when removed from the protective cover, all were visible enough to discard the product prior to use.There is no way to remove without contaminating the item.
 
Additional Manufacturer Narrative
Exemption number e2016018.B.Braun inc.(bbmi) (importer) is submitting the report on behalf of b.Braun melsungen ag (manufacturer).This report has been identified as event four of b.Braun medical inc.Internal report number (b)(4).One sample was received for evaluation.No abnormalities or defects were observed on the returned sample.Review of the device history record performed for the reported lot number did not reveal any abnormalities or nonconformance's of this nature.No adverse quality trends of this nature were identified during the complaint review process for the reported catalog number or lot number.Based on the results of this investigation, no specific conclusions can be made regarding the cause of the reported event.If additional pertinent information becomes available a follow-up report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTROCAN SAFETY®
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun-str. 1
melsungen, hessen 34212
GM  34212
MDR Report Key8427983
Report Number9610825-2019-00061
Device Sequence Number1341356
Product Code DYB
UDI-Device Identifier04046963185206
UDI-Public(01)04046963185206
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K021094
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/25/2019
Report Date (Section F) 04/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/01/2023
Device Model Number4251652-02
Device Catalogue Number4251652-02
Device Lot Number18H18G8391
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2019
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date02/19/2019
Device Age6 MO
Event Location Hospital
Date Report to Manufacturer04/25/2019
Initial Date Received by Manufacturer 02/19/2019
Supplement Date Received by Manufacturer02/19/2019
Initial Report FDA Received Date03/18/2019
Supplement Report FDA Received Date04/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA04/25/2019
-
-