• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD BIOSYN; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD BIOSYN; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID Back to Search Results
Model Number GM-322
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/21/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during open breast mastectomy, while closing subcuticular layer near the breast area, the two sutures came off from the swage after taking 2 or 3 bites.Another suture was used to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) received nine devices.A tactile and microscopic inspection of the sutures was performed with no abnormalities.The visual inspection of the opened sample noted three needles and two sutures.The needles were returned disengaged from the sutures.Under microscopic inspection, instrument marks were noted at the swaged end of two needles.A needle attachment test was performed on the sealed samples, and the results met the specifications for this product.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.Grasping the needle at the swaged end can cause the needle to break, bend, or become disengaged from the suture.The needle should always be grasped in the middle, where the diameter is greatest.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOSYN
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 18750
DO  18750
MDR Report Key8428233
Report Number9612501-2019-00538
Device Sequence Number1340767
Product Code GAM
UDI-Device Identifier10884521032699
UDI-Public10884521032699
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K000037
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2023
Device Model NumberGM-322
Device Catalogue NumberGM-322
Device Lot NumberD8F0870X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2019
Initial Date Received by Manufacturer 02/21/2019
Supplement Date Received by Manufacturer03/19/2019
Initial Report FDA Received Date03/18/2019
Supplement Report FDA Received Date04/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-