• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEUWAVE MEDICAL, INC. NEUWAVE PR PROBE 20CM 17GA; SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEUWAVE MEDICAL, INC. NEUWAVE PR PROBE 20CM 17GA; SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES Back to Search Results
Catalog Number PR20
Health Effect - Clinical Codes Unspecified Infection (1930); Tissue Damage (2104); Hypoesthesia (2352)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).The lot was not provided; therefore, a manufacturing record evaluation could not be performed.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Does the physician believe that the neuwave device caused or contributed to the patient complications mentioned in the article? if yes, please explain.
 
Event or Problem Description
It was reported via literature entitled: microwave ablation of osteoid osteoma: initial experience and efficacy.Author(s): elliot s.Rinzler, giridhar m.Shivaram, dennis w.Shaw, eric j.Monroe, kevin s.H.Koo.Citation: https://doi.Org/10.1007/s00247-018-4327-1.The purpose of this study was to retrospectively evaluate the technical feasibility and clinical efficacy of microwave ablation in the treatment of pediatric osteoid osteomas.Between january 1, 2015, and may 31, 2018, 24 pediatric patients, age range: 3-18; male: 15, female: 9 were evaluated and included in this study.During the procedure, the coring needle was then removed and the 17-gauge neuwave pr antenna (ethicon us, llc, somerville, nj) was advanced just beyond the distal edge of the lesion to cover the entire nidus within the antenna¿s ablation zone.A minimum of three ablation cycles were performed at a power setting of 30 watts, with a target temperature of 90 degrees celsius and a 30-s cooling period between cycles.Each ablation cycle lasted 30 s.Minor complications were seen in 4/24 (17%) patients.These included 3 patients with local skin numbness in the region of the access site (class a complication, two of which had spontaneously resolved at their 3- month follow-up), and one patient with a soft-tissue infection at the access site (class b complication, resolved with antibiotic treatment).There were no major complications.In conclusion, this study demonstrates that microwave ablation is a safe and efficacious technique for treating pediatric osteoid osteomas.The theoretical advantages of this technique require further study to determine non-inferiority to other established techniques and to evaluate long-term clinical response.Specific areas of future research could include determining the specific benefit of faster ablative techniques on procedural and anesthetic duration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUWAVE PR PROBE 20CM 17GA
Common Device Name
SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES
Manufacturer (Section D)
NEUWAVE MEDICAL, INC.
3529 anderson st
madison 53704
MDR Report Key8428659
Report Number3008769756-2019-00020
Device Sequence Number222442
Product Code NEY
UDI-Device Identifier00853719006418
UDI-Public00853719006418
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K160936
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source literature
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberPR20
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/21/2019
Initial Report FDA Received Date03/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SYSTEM - NWA1US1N
Outcome Attributed to Adverse Event Required Intervention;
-
-