• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. MCK TIBIAL BASEPLATE-LM/RL-SZ 3; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. MCK TIBIAL BASEPLATE-LM/RL-SZ 3; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLY Back to Search Results
Catalog Number 180603
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Injury (2348)
Date of Event 02/18/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported that the patient's left partial knee was revised due to subsidence of the baseplate.A mako partial knee was revised to a triathlon total knee construct including patella.Rep provided primary and revision implant sheets as well as x-rays and reported that no further information is available.
 
Event or Problem Description
It was reported that the patient's left partial knee was revised due to subsidence of the baseplate.A mako partial knee was revised to a triathlon total knee construct including patella.Rep provided primary and revision implant sheets as well as x-rays and reported that no further information is available.
 
Additional Manufacturer Narrative
An event regarding subsidence involving a mako baseplate was reported.The event was confirmed by medical review.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Medical records received and evaluation: a review of the provided medical records by a clinical consultant stated the following comment: x-ray confirms subsided tibial component, need operative reports, clinical and past medical history and serial x-rays.Product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, operative reports, clinical and past medical history and serial x-rays are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCK TIBIAL BASEPLATE-LM/RL-SZ 3
Common Device Name
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLY
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key8429285
Report Number3005985723-2019-00240
Device Sequence Number2989345
Product Code NPJ
UDI-Device Identifier00848486000691
UDI-Public00848486000691
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K090763
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Catalogue Number180603
Device Lot Number26221017-01
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/18/2019
Supplement Date Received by Manufacturer05/08/2019
Initial Report FDA Received Date03/18/2019
Supplement Report FDA Received Date06/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age73 YR
-
-