| Model Number |
37304 |
| Medical Device Problem Code |
Nonstandard Device (1420)
|
| Health Effect - Clinical Code |
Urinary Tract Infection (2120)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the male external catheter was too narrow where it connects to the patient tubing; and allows urine to flow back into the catheter.The patients alleged that uti (urinary tract infections) have occurred during the use of the male external catheter.It is unknown at this time if treatment was received for the urinary tract infections.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed.The visual evaluation noted 33 unopened spirit mecs were received.No obvious defects.The samples were tested according to the general inspection level ii/single sampling plan.Eight samples were opened and the funnel inner diameters were measured.The id's measured to be 0.3215", 0.3260", 0.3320", 0.3240", 0.3285", 0.3240", 0.3345", 0.3365", which were found to be within specification (0.33" +/- 0.01").The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "description/indication the self-adhering male external catheter is designed for the management of male urinary incontinence.Contraindication do not use on irritated or compromised skin.Precaution do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply wash penis with mild soap and warm water.Dry thoroughly.Trim pubic hair if necessary.Open package at perforation.To remove plastic insert, squeeze catheter at the top of the white cone and pull to release.Unroll self-adhering catheter over penis.Gently squeeze the catheter to properly seal adhesive to the skin.Connect to collection bag.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive.".
|
| |
|
Event or Problem Description
|
|
It was reported that the male external catheter was too narrow where it connects to the patient tubing; and allows urine to flow back into the catheter.The patients alleged that uti (urinary tract infections) have occurred during the use of the male external catheter.It is unknown at this time if treatment was received for the urinary tract infections.
|
| |
|