Catalog Number 111140
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 02/18/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Not available.
 
Event or Problem Description
The tip of the slap hammer broke off inside the trial femur.
 
Additional Manufacturer Narrative
An event regarding crack/fracture involving a mako slap hammer was reported.The event was confirmed by visual inspection.Method & results: -device evaluation and results: visual inspection: the reported device was not returned however a photograph was provided for review.From the photograph provided there is evidence that the slap hammer has fractured and that the fractured piece is stuck in the femoral trial component.Material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Even if the device was returned, the device is fractured and in its current condition would not be an accurate reflection of its original manufactured condition.-clinician review: no medical records were received for review with a clinical consultant.-device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there has been one other similar event for the reported lot.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re- opened.
 
Event or Problem Description
The tip of the slap hammer broke off inside the trial femur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL TRIAL SLAPHAMMER
Common Device Name
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key8429996
Report Number3005985723-2019-00243
Device Sequence Number16804226
Product Code NPJ
UDI-Device Identifier00848486002732
UDI-Public00848486002732
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K090763
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number111140
Device Lot Number06050317
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/18/2019
Supplement Date Received by Manufacturer10/21/2019
Initial Report FDA Received Date03/18/2019
Supplement Report FDA Received Date11/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age60 YR
-
-