• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR SILTEX ROUND SPECTRUM; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR SILTEX ROUND SPECTRUM; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Catalog Number 3542440M
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 02/19/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed for the finished device number 84477, and no non-conformances related to the reported complaint condition were identified.Concomitant products: mentor siltex round spectrum 275cc prosthesis, catalog #3542440m, lot #96815.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) year-old caucasian female underwent breast reconstruction revision with a mentor siltex round spectrum 275cc prosthesis and experienced right sided deflation post procedure.The deflation was diagnosed by a physician.As a result, an explant is planned for (b)(6) 2019.
 
Additional Manufacturer Narrative
The investigation of the returned device was completed by the failure analysis lab on 6/12/2019.Device evaluation summary: according to the information provided, it was reported that the patient experienced a deflation of the implant.During evaluation of the sample, it appeared intact.Leak testing was performed, and it revealed a tear measuring approximately 0.4 cm in the posterior view.Microscopic examination was performed and no evidence of instrument damage was observed.No other anomalies were observed.Possible causes of deflation of saline-filled implants include, but are not limited to, the following events: excessive stresses or manipulations as may occur during normal daily routines including customary and purposeful trauma to the breast.The reported complaint was confirmed.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Based on the information reported and/or the product investigation, there is no evidence that the issue is related with the manufacturing process.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SILTEX ROUND SPECTRUM
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key8430726
Report Number1645337-2019-09431
Device Sequence Number13502323
Product Code FWM
UDI-Device Identifier00081317002086
UDI-Public00081317002086
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year1995
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3542440M
Device Lot Number84477
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2019
Initial Date Received by Manufacturer 02/25/2019
Supplement Date Received by Manufacturer05/28/2019
Initial Report FDA Received Date03/18/2019
Supplement Report FDA Received Date06/24/2019
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
-
-