| Model Number |
EGIATRS45AMT |
| Medical Device Problem Codes |
Misfire (2532); Failure to Form Staple (2579)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/22/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic video-assisted thoracoscopic surgery bullectomy, when it was being applied to the lung to take the bulla out, when they started to fire, there was a loud crack and they were not able to compress it again.The staple line was incomplete distally.They opened a second handle and the same thing happened with the three reloads and then with the other two types of reloads as well.They opened another stapler and was fired at a different angle to the existing firings and it worked to complete the case.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the returned product noted that the reload was partially fired.Staple pushers were visible at the cut line.Microscopic evaluation noted that the reload had damage to the cutting edge of the knife blade.The clamping mechanism was deformed.Functionally the reload was loaded into a post market vigilance instrument, the interlock was overridden, and the reload was applied to test media.All staples were placed, and test media was cleanly transected.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.Replication of the damaged knife blade may occur when an obstacle has been incorporated in the jaws during application.The information booklet which accompanies each product shipment cautions the user; ensure that no obstructions (such as clips) are incorporated in the instrument jaws.Firing over an obstruction may result in incomplete cutting action and/or improperly formed staples.Replication of the bowed anvil, deformed anvil clamping mechanism and partially flushed staple pushers.Application over tissue that is beyond the recommended thickness range.Application with an obstacle incorporated in the jaws.In any of these circumstances, it will become increasingly difficult to actuate the firing handle and the instrument return knobs will be difficult to retract.In addition, staples may not form properly and tissue may not be fully transected.The information booklet which accompanies each product shipment offers the following as a warning and precaution.Preoperative radiotherapy may result in changes to tissue.These changes may, for example, cause the tissue thickness to exceed the indicated range thickness for the staple size.Careful consideration should be given to any pre-surgical treatment the patient may have undergone and in corresponding selection of staple size.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic video-assisted thoracoscopic surgery bullectomy, when it was being applied on the lung to take the bulla out, when they started to fire, there was a loud crack and they were not able to compress it again.The device partially fired and the staple line was incomplete distally.They opened a second handle and same thing happened with the three reloads and then with the other two types of reloads as well.They opened another stapler and was fired at a different angle to the existing firings and it worked to complete the case.There was no patient injury.
|
| |
|