Model Number IBI-81102
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One decapolar, inquiry steerable diagnostic catheter was received for evaluation.The catheter was returned due to a deflection issue which was not confirmed.The catheter shaft deflected when actuating the steering mechanism and deflected in the correct shape according to specifications.An incidental finding that electrode 8, 9, and 10 were all in the wrong position was noted while inspecting the device.Investigation confirmed that the displacement of the 8, 9 and 10 electrodes is consistent to the manipulation of the catheter during use of the introducer sheath.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.
 
Event or Problem Description
This report is to advise of an event observed during analysis confirming a displaced electrode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INQUIRY¿ ELECTROPHYSIOLOGY CATHETERS, DECAPOLAR, 6F
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
MDR Report Key8432705
Report Number3008452825-2019-00114
Device Sequence Number13898406
Product Code DRF
UDI-Device Identifier05414734302968
UDI-Public05414734302968
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K961924
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2021
Device Model NumberIBI-81102
Device Catalogue NumberIBI-81102
Device Lot Number6600479
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2019
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2019
Initial Report FDA Received Date03/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No