Model Number TABLETOP |
Device Problems
Mechanical Problem (1384); Firing Problem (4011)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/22/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
|
|
Event Description
|
A customer reported that there was continuous shooting of the laser.The foot pedal locked in the lower position during a vitrectomy procedure.An alternate foot pedal was used to complete the procedure.There was no patient harm.Additional information has been requested.
|
|
Manufacturer Narrative
|
The company service representative examined the system and was able to confirm and replicate the footswitch issue of locking in the down position.The footswitch was replaced to resolve the issue between surgeries.The company service representative was not able to test the system per the service test procedure (stp).The system manufacturing device history record (dhr) was reviewed.Based on assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to the nonconforming laser footswitch.The manufacturer internal reference number is: (b)(4).
|
|