• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR FEM COMP #1 L-CEM; KNEE JOINT PATELLOFEMOROTIBIAL METAL/POLYMER POROUS-COATED UNCEMENTED PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR FEM COMP #1 L-CEM; KNEE JOINT PATELLOFEMOROTIBIAL METAL/POLYMER POROUS-COATED UNCEMENTED PROSTHESIS Back to Search Results
Catalog Number 5510F101
Medical Device Problem Codes Malposition of Device (2616); Positioning Problem (3009)
Health Effect - Clinical Codes Pain (1994); Swelling (2091); Injury (2348)
Date of Event 11/14/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An event regarding baseplate loosening was reported, malposition of the triathlon femoral component was confirmed based on clinician review of the x-ray provided.Method & results: -product evaluation and results: damage was observed on the femoral component and baseplate, consistent with contact against each other.Delamination, burnishing, scratches and third-body indentations were observed on the insert condyles.These are common damage modes of uhmwpe.Debris was observed on the lateral condyle of the insert.Backside impression markings were observed on distal surface of the insert, consistent with contact against the baseplate.Eds showed the femoral component, baseplate and debris from the insert were all consistent with astm f75 alloy.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.-clinician review: a clinician review of the provided medical records states the following; provided undated x-ray confirms subluxated femoral component posteriorly.Need operative reports, clinical reports, histopathology reports, office notes, serial x-rays and examination of the explanted components.- product history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: no other similar events were reported for the lot indicated.Conclusions: a clinician review of the provided medical records states the following; provided undated x-ray confirms subluxated femoral component posteriorly.Need operative reports, clinical reports, histopathology reports, office notes, serial x-rays and examination of the explanted components.The device was returned and material analysis was performed which confirmed damage was observed on the femoral component and baseplate, consistent with contact against each other.Delamination, burnishing, scratches and third-body indentations were observed on the insert condyles.These are common damage modes of uhmwpe.Debris was observed on the lateral condyle of the insert.Backside impression markings were observed on distal surface of the insert, consistent with contact against the baseplate.Eds showed the femoral component, baseplate and debris from the insert were all consistent with astm f75 alloy.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.The exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, clinical reports, histopathology reports, office notes and serial x-rays are needed to fully investigate the event.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported that patient's left knee was revised.As reported by rep: ".Progressive pain, swelling, sense of unreliability.Ct and x-rays revealed tibial loosening.Soft tissue effusion, cell count (3,500) and synovasure worrisome for possible infection.Cultures negative to date".Update: a clinician review of the provided medical records states the following; provided undated x-ray confirms subluxated femoral component posteriorly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR FEM COMP #1 L-CEM
Common Device Name
KNEE JOINT PATELLOFEMOROTIBIAL METAL/POLYMER POROUS-COATED UNCEMENTED PROSTHESIS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8433739
Report Number0002249697-2019-01368
Device Sequence Number1663850
Product Code MBH
UDI-Device Identifier07613327040203
UDI-Public07613327040203
Combination Product (Y/N)N
Initial Reporter StateVT
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2011
Device Catalogue Number5510F101
Device Lot NumberSCXEH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2019
Initial Report FDA Received Date03/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age75 YR
Patient Weight62
-
-