Catalog Number 657412S
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/20/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Upon completion of the investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
While driving a bone screw into a distal fibula plate, metal wires(chips) came from the screw.No surgical delay or adverse consequences were reported.Procedure was completed successfully.
 
Additional Manufacturer Narrative
The reported event could be confirmed.Based on investigation, the root cause was attributed to be user related.The failure was caused by possible inadequate angulation during screw insertion.The device inspection revealed the following: some metal shaving (coils) on both returned screws is clearly evident.The metal coils are still attached to the screw bodies.The remaining coil on the screw with lot y44299 is thin, whereas the coil on lot y36493 is rather thick.Also the abrasions on both screws are almost for the entire length to the second last thread turn.Which indicates that the positioning of the plate and the predrilled hole were initially not line up accordingly and the applied angle during screw insertion resulted in the screw abrasions (coil).It looks as the plate was repositioned but still not enough in order to insert the second screw as there were still some thinner abrasions evident.Finally new screw(s) was inserted without any further problems (possibly after realigning the plate).A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any further information is provided, the investigation report will be updated.
 
Event or Problem Description
While driving a bone screw into a distal fibula plate, metal wires(chips) came from the screw.No surgical delay or adverse consequences were reported.Procedure was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE SCREW VARIAX2 T10, FULL THREAD, 3.5MM / L12MM
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
MDR Report Key8434674
Report Number0008031020-2019-00276
Device Sequence Number13743903
Product Code HRS
UDI-Device Identifier07613327073768
UDI-Public07613327073768
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K132502
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number657412S
Device Lot NumberY44299
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2019
Initial Date Received by Manufacturer 02/20/2019
Supplement Date Received by Manufacturer03/26/2019
Initial Report FDA Received Date03/19/2019
Supplement Report FDA Received Date04/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-