• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JOURNEY TIBIA BASE NP LT SZ 7; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JOURNEY TIBIA BASE NP LT SZ 7; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY Back to Search Results
Catalog Number 74022227
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 11/12/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after underwent a total knee arthroplasty for a study, the patient had delayed wound healing and possible infection.Revision surgery and drainage were performed with a poly exchange.The subject recovered from the event but was terminated from the study.The explanted device was destroyed.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned.A clinical evaluation was conducted and x-rays provided do not contribute to the reported wound infection.Based on the limited information provide and without the return of the explanted device for analysis the root cause of the reported delayed healing cannot be confirmed at this time but the circumstances reported of the delayed wound healing is a known complication of a surgical procedure and not a failure of the implant devices.Based on the lab results there was no indication of infection.The patient¿s medical history is significant for multiple comorbidities i.E.Diabetes which cannot be ruled out as possible contributing factors.The future impact to the patient beyond the revision cannot be determined.No further clinical/medical assessment is warranted at this time.A review of complaint history on the listed parts revealed no prior complaints for the listed batches with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This included a review of sterilization documents which indicated that the product was sterilized according to sterilization release documentation.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY TIBIA BASE NP LT SZ 7
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8434687
Report Number1020279-2019-01115
Device Sequence Number1780950
Product Code JWH
UDI-Device Identifier03596010554406
UDI-Public03596010554406
Combination Product (Y/N)N
PMA/510(K) Number
K121443
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source health professional,study,use
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number74022227
Device Lot Number18DM09696
Initial Date Received by Manufacturer 02/26/2019
Supplement Date Received by Manufacturer02/26/2019
Initial Report FDA Received Date03/19/2019
Supplement Report FDA Received Date06/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age60 YR
-
-