Catalog Number 4007
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/22/2019
Event Type  malfunction  
Event Description
It was reported that when the carrier #4 was removed, the film dressing got peeled off from the skin due to the strong adhesion between the film and the carrier #4.No patient harm.No delay.Backup dressings were available.
 
Manufacturer Narrative
We have now concluded our investigation into this complaint.The batch records were reviewed and it could be confirmed that the products had progressed through a satisfactory release process which involved testing the product against the requirements of its finished product specification.There was no issue identified during the manufacturing process that could have caused or contributed to the complaint problem.Complaint samples were received and assessed.They were pasted onto the arm and the non adhesive tab at the end of the dressing was lifted to peel the carrier from the dressing.The carrier was hardly separated from the soft film and some of the samples showed carrier delamination and a thin layer left on the film.A review of the complaint history for this defect shows a very low level of complaints for this issue based on volumes manufactured.The results of the investigation have shown that the most probable cause was an incorrect knife temperature setting.This has now been adjusted and the in process inspection method optimised for carrier peeling.The delamination will also be raised with the raw material supplier.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.We will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IV3000 1 HAND 6X7CM CTN 100
Type of Device
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull
UK 
MDR Report Key8438252
MDR Text Key139380787
Report Number8043484-2019-00198
Device Sequence Number1
Product Code KGX
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/20/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number4007
Device Lot Number201831
Date Manufacturer Received04/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-