| Model Number |
48-50-00 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/07/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Livanova (b)(4) manufactures the s5 system.The incident occurred in (b)(4).A livanova representative was able to confirm the reported issue.The technician was able to trace the failure back to a defective board.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Livanova (b)(4) received a report that a burning smell was noticed during the use of a s5 system.It was further reported that after the burning smell, an error message was displayed on the s5 system.There was no report of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The board was returned to livanova deutschland for further investigation.During the evaluation the reported issue could be confirmed and reproduced.It was found that a defective component on the bard led to the reported malfunction.
|
| |
|