| Catalog Number |
365987 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that during use of the bd microtainer® pst¿ tubes with lh (lithium heparin) there was an issue of leakage.There were two additional reports of leakage.Since these two were reported we have had 2 more incidents that i am aware of.One on feb 9th and one on feb 10th.Feb 9th ¿ only information i have is that it was a baby and the sample was recollected.Tube was not saved.Feb 10th ¿ patient was an (b)(6) year old female from emergency department.Sample was analyzed for ck and troponin on roche cobas.Results were reported.[ck ¿ 29 (range 28-110), htnt ¿ 18 (range <14)] empty tube is available.Both of these would have also resulted in exposure and cleanup of blood spill in centrifuge.All of these samples were spun in the same centrifuge but each one was in a different position.All were the same lot # 827652n.There have been no more incidents that i am aware of since feb 10th.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation: investigation summary: bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for a molding issue on the tube bottom of the reservoir, was observed.The molding issue likely attributed to the leakage that the customer had observed with the incident lot.Additionally, retention samples were selected from bd inventory for evaluation and upon completion, no issues were found relating to this issue.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Bd had initiated further investigation relating to this issue through capa#766109 and the potential causes were identified.As a result, corrective actions were established and implemented.Investigation conclusion: based on evaluation of the customer samples, the customer¿s indicated failure mode for a molding issue on the tube bottom of the reservoir was observed.Evaluation of the retain samples was also conducted and no issues were observed.Further investigation activities were conducted through capa#766109 and the potential root causes were identified.As a result, corrective actions and procedures were implemented to mitigate further occurrences.Root cause description: capa#766109 was conducted to document further investigation and root cause analysis relating to this issue.The investigation identifed the most likely root causes and corrective actions were implemented.Rationale: based on an assessment of severity and frequency, it was determined that a capa was required in order to determine the root cause associated with this issue.The investigation identified potential root causes for this issue and corrective actions were implemented.
|
| |
|
Event or Problem Description
|
|
It was reported that during use of the bd microtainer® pst¿ tubes with lh (lithium heparin) there was an issue of leakage.There were two additional reports of leakage since these two were reported we have had 2 more incidents that i am aware of.One on feb 9th and one on feb 10th.Feb 9th ¿ only information i have is that it was a baby and the sample was recollected.Tube was not saved.Feb 10th ¿ patient was an 87 year old female from emergency department.Sample was analyzed for ck and troponin on roche cobas.Results were reported.[ck ¿ 29 (range 28-110), htnt ¿ 18 (range <14)] empty tube is available.Both of these would have also resulted in exposure and cleanup of blood spill in centrifuge.All of these samples were spun in the same centrifuge but each one was in a different position.All were the same lot # 827652n.There have been no more incidents that i am aware of since feb 10th.
|
| |
|
Event or Problem Description
|
|
It was reported that during use of the bd microtainer® pst¿ tubes with lh (lithium heparin) there was an issue of leakage.There were two additional reports of leakage since these two were reported we have had 2 more incidents that i am aware of.One on feb 9th and one on feb 10th.Feb 9th: only information i have is that it was a baby and the sample was recollected.Tube was not saved.Feb 10th: patient was an 87 year old female from emergency department.Sample was analyzed for ck and troponin on roche cobas.Results were reported.[ck ¿ 29 (range 28-110), htnt ¿ 18 (range <14)].Empty tube is available.Both of these would have also resulted in exposure and cleanup of blood spill in centrifuge.All of these samples were spun in the same centrifuge but each one was in a different position.All were the same lot # 827652n.There have been no more incidents that i am aware of since feb 10th.
|
| |
|
Additional Manufacturer Narrative
|
|
Bd is conducting a voluntary medical device recall for the previously reported catalog and lot numbers for the bd microtainer® tubes based on confirmation that there may be damaged tube reservoirs present.A damaged reservoir may lead to a decreased fill volume causing samples to be insufficient for testing and improper blood-to-additive ratio, potentially producing erroneous results. recollecting samples and retesting for the patient may be required if a microtainer® tube with a damaged reservoir was used.This may lead to a delay in test results being reported and patients being treated. additionally, a damaged tube reservoir may result in the potential for blood leakage and exposure to healthcare workers.The root cause investigation has determined that the damage to the tube reservoirs was caused by an equipment malfunction during the molding process. the equipment malfunction has been corrected. please reference bd recall #: pas-19-1454-fa.
|
| |
|