ABBOTT DIABETES CARE INC FREESTYLE LIBRE PRO; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71562-01
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Purulent Discharge (1812); Erythema (1840); Unspecified Infection (1930); Skin Irritation (2076); Swelling (2091)
Date of Event 02/18/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
An hcp reported a customer experienced an allergic skin reaction on day 3 of wear of the adc freestyle libre pro sensor.The customer presented with "redness and bumps" just on her right arm and self-treated with benadryl, claritin and hydrocortisone cream.A couple of days later, the redness and bumps from the right upper arm had spread and "weeping yellowish" discharge was noted.The customer had hcp contact and was administered an injection of the steroid, kenalog (dose unknown,) and was given a prescription for augmentin (amoxicillin, dose unknown) for treatment.There was no report of death or permanent injury associated with this event.
 
Additional Manufacturer Narrative
No product has been returned.Extended investigation has been performed for the reported complaint there was no indication that the product did not meet specification.Dhrs (device history review) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and freestyle libre sensor kit passed all tests prior to release.Dose audit reports were reviewed and demonstrates the continued effectiveness of the established sterilization process for libre sensor kits.Environmental monitoring reports were reviewed, including bioburden and endotoxin testing, and demonstrated that all monitoring processes continue to meet adc minimum requirements for product quality.All review activities conducted above, including but not limited to the final release testing specifically associated with the manufacture of this product, are sufficient information in order to show if the product has met specifications prior to release.If the product is returned, the case will be re-opened and a physical investigation will be performed.
 
Event or Problem Description
An hcp reported a customer experienced an allergic skin reaction on day 3 of wear of the adc freestyle libre pro sensor.The customer presented with "redness and bumps" just on her right arm and self-treated with benadryl, claritin and hydrocortisone cream.A couple of days later, the redness and bumps from the right upper arm had spread and "weeping yellowish" discharge was noted.The customer had hcp contact and was administered an injection of the steroid, kenalog (dose unknown,) and was given a prescription for augmentin (amoxicillin, dose unknown) for treatment.There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE PRO
Common Device Name
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
MDR Report Key8440373
Report Number2954323-2019-02333
Device Sequence Number1104595
Product Code MDS
UDI-Device Identifier00357599715625
UDI-Public00357599715625
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
P150021
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number71562-01
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/19/2019
Supplement Date Received by Manufacturer05/21/2019
Initial Report FDA Received Date03/21/2019
Supplement Report FDA Received Date06/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Weight120
-
-