|
Model number/catalog number: sc-3400-30, serial number: (b)(4), batch/lot number: 15658262, model/catalog description: splitter 2x8 kit-sterile.The explanted devices were not returned to bsn as they were discarded by the medical facility.It is indicated that the devices will not be returned for evaluation; therefore a failure analysis of the complaint devices could not be completed.A review of the device history records will be conducted.If there is any further relevant information from that review, a supplemental med watch will be filed.
|
|
A report was received that there were high impedances noted in the patient's scs system.The patient underwent a revision procedure wherein the splitters were replaced.There were still high impedances, however, all pain areas were covered.The patient was doing well postoperatively.
|