| Model Number |
7010 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/04/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older.Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed that the hypotube is separated 70.8cm from the hub.The separation appears to have been kinked prior to separation.Microscopic examination revealed no additional damages.There is blood present in the guidewire lumen and on the balloon folds.The balloon is still tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.There was no evidence of any damage or irregularities contributing to the reported crossing difficulty, which could not be confirmed because the clinical circumstances could not be replicated.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on (b)(6) 2019.It was reported that crossing difficulties were encountered.The 90% stenosed target lesion was located in a severely tortuous and severely calcified left anterior descending artery.A 3.75mm x 12mm quantum maverick balloon catheter was advanced but failed to cross the lesion.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed separated hypotube.
|
| |
|
Search Alerts/Recalls
|