• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS, INC. ADVIA CENTAUR XP TNI-ULTRA ASSAY; TROPONIN IMMUNOASSAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS, INC. ADVIA CENTAUR XP TNI-ULTRA ASSAY; TROPONIN IMMUNOASSAY Back to Search Results
Model Number N/A
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/27/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Siemens reviewed the information provided by the customer.The quality control (qc) was in range at the time the sample was run.The customer ran the sample on a heterophile binding tube (hbt) tube - this did not decrease the result, it remained elevated indicating that a heterophile antibody is not present in the sample.There was not sufficient sample to send to siemens for any further testing.Based on information provided, an interfering substance in the sample cannot be ruled out, causing the elevated result on the advia centaur xp and not on the dimension exl.The dimension troponin i method uses different antibodies and has different specificity therefore may not be susceptible to the same interference seen with this sample.Immunoassays are subject to a number of interferences including those caused by endogenous antibodies and also other exogenous interferences such as drugs, nutritional supplements and/or herbal medicine in the blood.This issue was isolated to this patient sample.Based on the investigation, no product problem was identified.The instrument is performing within specifications.No further evaluation of the device is required.The instruction for use (ifu) under the summary and explanation of the test section states the following: "always analyze ctni results in the context of time elapsed since patient presentation to the hospital.Serial sampling is recommended to detect the temporal rise and fall of troponin levels characteristic of mi.In accord with published recommendations, serial testing of ctni at intervals of 2 to 4 hours for up to 12 to 24 hours is suggested to corroborate a single ctni result.An elevated troponin alone is not sufficient to make the diagnosis of mi.Other markers, such as ck-mb and myoglobin, can be used in conjunction with ctni results in aiding the diagnosis of mi." mdr 1219913-2019-00042 was filed for the same event (sid (b)(6)).
 
Event or Problem Description
A false positive advia centaur xp troponin ultra (tni-ultra) result was obtained for a patient sample on (b)(6) 2019.The patient sample was repeated and the result was positive.The positive result was reported to the physician and questioned.The physician sent the patient to (b)(6) hospital where another sample was obtained and tested for troponin on an alternate method.The result was negative.The patient had a sample collected on (b)(6) 2019 and tested on the advia centaur xp for tni-ultra.The result was positive.The sample was repeated and the result was positive.The result was reported to the physician and was questioned.The sample was sent internally to a branch laboratory for analysis on the siemens dimension exl and the result was negative.A corrected report was issued.The patient did not have any clinical features of a cardiac event.Patient treatment was not prescribed or altered.There was no report of adverse health consequences due to the discordant tni-ultra result.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA CENTAUR XP TNI-ULTRA ASSAY
Common Device Name
TROPONIN IMMUNOASSAY
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
511 benedict avenue
tarrytown NY 10591 5097
MDR Report Key8441528
Report Number1219913-2019-00041
Device Sequence Number11207813
Product Code MMI
UDI-Device Identifier00630414516226
UDI-Public00630414516226
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K053020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2019
Device Model NumberN/A
Device Catalogue Number10317709
Device Lot Number54907139
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2019
Initial Report FDA Received Date03/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age51 YR
-
-