Model Number MMT-523LNAS
Medical Device Problem Codes Connection Problem (2900); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/15/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
The customer reported via phone call that insulin pump had hipped and cracked around reservoir area o ring has come out and mentioned does not feel reservoir is secure.The blood glucose at the time of the incident was unknown.The customer was assisted with troubleshooting.The device will not be returned for analysis.
 
Additional Manufacturer Narrative
Insulin pump was received with a broken reservoir tube lip and missing reservoir tube o-ring.A test reservoir was installed and did not lock in place and unable to perform the displacement test due to broken reservoir tube lip and missing reservoir tube o-ring.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAS
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8442556
Report Number3004209178-2019-70247
Device Sequence Number1796443
Product Code OYC
UDI-Device Identifier00643169479630
UDI-Public(01)00643169479630
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523LNAS
Device Catalogue NumberMMT-523LNAS
Device Lot NumberA3523LNASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/13/2019
Initial Date Received by Manufacturer 03/15/2019
Supplement Date Received by Manufacturer11/05/2019
Initial Report FDA Received Date03/22/2019
Supplement Report FDA Received Date11/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
-
-