| Model Number |
MMT-523LNAS |
| Medical Device Problem Codes |
Connection Problem (2900); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that insulin pump had hipped and cracked around reservoir area o ring has come out and mentioned does not feel reservoir is secure.The blood glucose at the time of the incident was unknown.The customer was assisted with troubleshooting.The device will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump was received with a broken reservoir tube lip and missing reservoir tube o-ring.A test reservoir was installed and did not lock in place and unable to perform the displacement test due to broken reservoir tube lip and missing reservoir tube o-ring.(b)(4).
|
| |
|