| Model Number |
MMT-523RNAL |
| Medical Device Problem Code |
Failure to Cycle (1142)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/23/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported that their insulin pump alarmed compromised force sensor system.The customer¿s blood glucose level was unknown.The customer also reported that the insulin was squirting out during manual prime process.The customer stated that the drive support cap was normal.The customer was advised to the insulin pump will need to be replaced and discontinue use of the insulin pump and to revert to the backup plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump passed the displacement test but compromised forced sensor system alarm during the basic occlusion test due to loose drive support disk.Unable to perform the occlusion, prime and excessive no delivery test due to compromised forced sensor system alarm.
|
| |
|