| Model Number |
60-02-15 |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
02/22/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
|
| |
|
Event or Problem Description
|
|
Livanova (b)(4) received a report that the encoder of a centrifugal pump system with tubing clamp would not control the flow speed.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Replaced encoder (b)(4) and tested unit.All functions works properly.Perfusion was satisfied with the repair and returned the unit back into service.A review of the dhr did not identify any deviations or nonconformities relevant to the issue.
|
| |
|
Event or Problem Description
|
|
See initial report.
|
| |
|