| Catalog Number |
2D72LS75 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/22/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The complaint was forwarded to the manufacturing facility for investigation.The device history record of the complaint lot was reviewed.The lot was inspected and released in compliance with all requirements.The sample is enroute to the manufacturing facility.Once their investigation of it is completed, a follow up report will be filed.
|
| |
|
Event or Problem Description
|
|
Cases were selective coronary angiograms (sca) with percutaneous coronary interventions (pci).During the pci the stickiness of the gloves would cause the coronary wire to dislodge from the lesion resulting in the physician needing to rewire the lesion.Per customer, no adverse effects to patients other than lengthened procedure times.No additional patient or event specifics provided by customer.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned samples were from an unopened package.They were evaluated by donning testing, and were found to be a little tacky.Grip testing was found to be within range.The device history record of the complaint lot was reviewed.The lot was inspected and released in compliance with all requirements.The manufacturing location had been moved, but the gloves continue to be manufactured using the same mold design and product dimensions, using the same test methods, and the same product specifications/quality standard.The chlorine concentration has been increased at the chlorination process to improve grip variation.We will continue to monitor complaints for any such issues.
|
| |
|