• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRN BCS II BOX PREP GD SZ3-5; PROSTHESIS KNEE PATELLOFEMOROTIBIAL SEMICONSTRAINED CEMENTED POLYMERMETALPOLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRN BCS II BOX PREP GD SZ3-5; PROSTHESIS KNEE PATELLOFEMOROTIBIAL SEMICONSTRAINED CEMENTED POLYMERMETALPOLYMER Back to Search Results
Catalog Number 74012575
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Information (3190)
Date of Event 02/27/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery item broke while it was being used.Screw fell out.Surgery was completed with same device.
 
Additional Manufacturer Narrative
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.One of the threaded pins in the base of the device was missing, rendering the device inoperable.The device was manufactured in 2014 and exhibits signs of extensive wear usage.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRN BCS II BOX PREP GD SZ3-5
Common Device Name
PROSTHESIS KNEE PATELLOFEMOROTIBIAL SEMICONSTRAINED CEMENTED POLYMERMETALPOLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8443980
Report Number1020279-2019-01145
Device Sequence Number19375678
Product Code JWH
UDI-Device Identifier00885556000991
UDI-Public00885556000991
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K111711
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number74012575
Device Lot NumberUNKNOWN
Initial Date Received by Manufacturer 02/27/2019
Supplement Date Received by Manufacturer02/27/2019
Initial Report FDA Received Date03/22/2019
Supplement Report FDA Received Date04/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-