• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MOUNTAIN HOME; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MOUNTAIN HOME; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number ASKU
Device Problems Fluid/Blood Leak (1250); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/24/2019
Event Type  malfunction  
Manufacturer Narrative
Use errors and proper user instructions are addressed in ¿the homechoice and homechoice pro systems patient at-home guide¿, which is shipped with every homechoice device.The guide warns the user not to perform dialysis in the same room as pets or animals.It states that ¿contamination of the fluid or fluid pathways can result if a pet or animal bites a solution bag or the disposable set.Contamination of any portion of the fluid or fluid path may result in peritonitis, serious patient injury, or death.¿ a review of the label for the product family will be conducted.The device was not returned and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a homechoice cassette leaked; further described as ¿fluid leakage from the cassette supply line tubing¿.The patient¿s cat came into the room and clawed at the tubing, and put a hole in the supply bag tubing, which caused the leak.This was identified during dwell one of five of automated peritoneal dialysis therapy.The patient was connected at the time of the event.Renal therapy services (rts) assisted the patient in ending therapy and with cassette removal.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NI
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
mountain home AR
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8444850
MDR Text Key139598831
Report Number1416980-2019-01534
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial
Report Date 03/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Date Manufacturer Received02/24/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
HOMECHOICE DEVICE
-
-