| Catalog Number |
824G |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/08/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).To date, the device has not been returned.If the product is returned for evaluation, any further information derived from the evaluation will be submitted in a supplemental 3500a form.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a mastopexy with prosthesis procedure on (b)(6) 2019 and suture was used.It was reported that the threads were bursting easily.A similar product was used to continue the procedure.No adverse patient consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Additional evaluation summary: an empty opened overwrap, a foil with an empty folder and a dispensed needle suture of product code 824, lot mbh211 were received for analysis.During the visual inspection of the sample, no defects were observed.The swage and attachment area were noted to be as expected.The suture was examined along of strand and no defects or damaged were noted.A functional test was performed on the samples and the tensile force met the requirements.Per the condition of the sample, no performance breakage suture was observed, and the tested sample met the finished goods requirements.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.
|
| |
|