• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. INTRODUCER KIT FOR GASTROSTOMY FEEDING TUBE, 22FR DILATOR; DH EF PERC PLACEMENT PRODUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. INTRODUCER KIT FOR GASTROSTOMY FEEDING TUBE, 22FR DILATOR; DH EF PERC PLACEMENT PRODUCTS Back to Search Results
Model Number 98433
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Peritonitis (2252)
Date of Event 02/20/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.Root cause could not be determined.The device history record for lot aa8029r05 was reviewed and the product was produced according to product specifications.All information reasonably known as of 21 mar 2019 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).
 
Event or Problem Description
It was reported that the anchor sutures for the gastropexy broke two hours after placement.The hospital alleges that as a result of the breaks, the patient developed peritonitis and required surgical intervention.Per additional information received 20 mar 2019, the patient's surgical intervention was successful and the patient is "safe.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRODUCER KIT FOR GASTROSTOMY FEEDING TUBE, 22FR DILATOR
Common Device Name
DH EF PERC PLACEMENT PRODUCTS
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
MDR Report Key8446364
Report Number9611594-2019-00056
Device Sequence Number429637
Product Code KGC
UDI-Device Identifier10680651984330
UDI-Public10680651984330
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K080253
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/01/2020
Device Model Number98433
Device Catalogue Number991098433
Device Lot NumberAA8029R05
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2019
Initial Report FDA Received Date03/22/2019
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-