MAUDE Adverse Event Report: SMITH & NEPHEW, INC. 40MM OXINIUM MODULAR HEAD; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. 40MM OXINIUM MODULAR HEAD; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Catalog Number 71342340
Medical Device Problem Code Unstable (1667)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 06/04/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that revision surgery was performed due to left anterior hip dislocation with recurrent instability.
 
Additional Manufacturer Narrative
The associated complaint device was not returned.The clinical/medical team concluded, the intraoperative findings of reddish colored fluid, synovitis, metallosis and bone loss are re consistent with findings associated with metal debris.Without the supporting analysis of the explanted components, the source of the reported metallosis cannot be confirmed, and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.Additionally, the patient¿s morbid obesity and the use of a zimmer acetabular component with a smith + nephew implant, which is contraindicated, and cannot be ruled out as a contributing factors for the dislocation and subsequent 2nd lt tha revision.It cannot be concluded that the reported reactions/events were associated with a mal-performance of the smith + nephew implant.The patient impact beyond the revision pain/healing cannot be determined.No further clinical assessment is warranted at this time.A review of the production documentation for the corresponding products did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed part revealed no prior complaints for the listed batch.Without the actual device involved our investigation cannot proceed.If the device or new information is received in the future, the complaints can be re-opened.No further actions are being taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
40MM OXINIUM MODULAR HEAD
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8446379
Report Number1020279-2019-01153
Device Sequence Number277430
Product Code MBL
UDI-Device Identifier00885556112410
UDI-Public00885556112410
Combination Product (Y/N)N
PMA/510(K) Number
K093363
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/06/2024
Device Catalogue Number71342340
Device Lot Number14DM07685
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/04/2019
Supplement Date Received by Manufacturer03/04/2019
Initial Report FDA Received Date03/22/2019
Supplement Report FDA Received Date10/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age53 YR
-
-