|
Dried blood could be found in the returned device.There were severely stretched and damaged were found in the (proximal and distal) shaft of teleport microcatheter, measured the stretched catheter (proximal and distal) shaft and confirmed that it has been stretched approximately 11.5cm.Microscopic examination revealed that the distal tip of the teleport microcatheter was still intact, but has been deformed and slightly damaged with flaring displayed.Damaged catheter outer body and deformed coil and exposed braided wire were found on the catheter shaft with significant signs of stretching, and the failure mode/appearance is consistent with damage of being trapped by some hard objects and rotation.Apart from above observed damage, no other anomalies were found on the returned device.An associated clinical picture was provided for analysis and the picture appears to show that the proximal shaft / distal shaft / distal tip of the microcatheter become stretched and damaged.A review of the clinical information for this complaint indicated that a catheter (teleport 2.0f x 150cm / lot# 4307271811) could not cross the lesion and became stuck at the same location as the previous device.Additional force was applied to remove the microcatheter and the microcatheter fractured and stretched during removal.The procedure was completed with angioplasty and atherectomy.During the manufacturing processes each microcatheter is inserted to a 0.0155" x 1800mm mandrel to inspect and ensure without obvious friction prior to proceed to next process.A review of the manufacturing records for this batch (4307271811) did not reveal any non-conformity which could have contributed to the complaint.A lot history review of the microcatheter was conducted and no manufacturing related causes could be found for this complaint.Orbusneich medical manufacturing processes include extensive testing and inspections to ensure each product meets all material, assembly, and performance specifications prior to release.The returned teleport microcatheter was examined in the failure analysis laboratory.The analysis results are: the damage on the teleport microcatheter was not due to a manufacturing assembling failure.The failure mode/appearance of the teleport microcatheter is consistent damage due to trapped with an external hard object (severe calcification) or other devices (such as associated guidewire) and further elongated by excessive pulling force.There are no signs of manufacturing defects on the returned product and / or the lot information from this product.Based on the above investigation, the possible root cause identified is operational context as device performance was limited due to anatomical procedure factors.No remedial action is being performed by orbusneich in response to this event at this time.
|