• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD REYNOSA S.A. DE C.V. EQUISTREAM CATHETER - STRAIGHT, POLYURETHANE CATHETER, STANDARD KIT WITH AIRGUA; DIALYSIS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD REYNOSA S.A. DE C.V. EQUISTREAM CATHETER - STRAIGHT, POLYURETHANE CATHETER, STANDARD KIT WITH AIRGUA; DIALYSIS CATHETER Back to Search Results
Model Number 5903270
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/28/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.
 
Event or Problem Description
It was reported that during dialysis catheter insertion, the tunneler component allegedly broke inside catheter and was removed.It was further reported that another catheter was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process, and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was not returned for evaluation.Images and medical records were not provided for review.Therefore, the investigation is inconclusive for the alleged broken tunneler as no objective evidence has been provided to confirm any alleged deficiency with the tunneler.The root cause could not be determined based upon available information.It is unknown whether patient and/or procedural factors contributed to the event.Possible contributing factors include excessive force and damaged prior to use; however, the lack of a returned sample prevented both confirmation of the reported event and identification of a root cause(s).Labeling review: a review of product labeling documents (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) showed that the product labeling is adequate.
 
Event or Problem Description
It was reported that during dialysis catheter insertion, the tunneler component allegedly broke inside catheter and was removed.It was further reported that another catheter was used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUISTREAM CATHETER - STRAIGHT, POLYURETHANE CATHETER, STANDARD KIT WITH AIRGUA
Common Device Name
DIALYSIS CATHETER
Manufacturer (Section D)
BARD REYNOSA S.A. DE C.V.
blvd montebello #1
parque industrial colonial
reynosa tamaulipas 88780
MX  88780
MDR Report Key8449076
Report Number3006260740-2019-00659
Device Sequence Number18918626
Product Code MSD
UDI-Device Identifier00801741013713
UDI-Public(01)00801741013713
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K090101
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5903270
Device Catalogue Number5903230
Device Lot NumberRECX3053
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/28/2019
Supplement Date Received by Manufacturer04/27/2019
Initial Report FDA Received Date03/25/2019
Supplement Report FDA Received Date05/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age50 YR
-
-