BOSTON SCIENTIFIC CORPORATION INNOVA VASCULAR; STENT, SUPERFICIAL FEMORAL ARTERY Back to Search Results
Model Number 26924
Medical Device Problem Codes Material Integrity Problem (2978); Activation Problem (4042)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age at time of event: 18 years or older.
 
Event or Problem Description
It was reported that stent damage occurred.A 6x40x130 innova vascular stent was selected for a percutaneous transluminal angioplasty procedure.The severely calcified, moderately tortuous, 99% stenosed lesion was accessed by ipsilateral approach.After the unspecified guidewire was passed through, and the stent delivery system was delivered to the target lesion, the physician tried to perform stenting with this device, but the thumbwheel would not move/activate while attempting to deploy the stent, so the device was removed.When checking the device after removal, the distal tip of the stent struts were found to be flared and lifted up.Therefore the device was replaced with an eluvia stent to complete the stenting.The same unspecified guidewire was used to complete the procedure.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOVA VASCULAR
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8449848
Report Number2134265-2019-01896
Device Sequence Number429591
Product Code NIP
UDI-Device Identifier08714729873990
UDI-Public08714729873990
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P140028
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/08/2020
Device Model Number26924
Device Catalogue Number26924
Device Lot Number0021231180
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2019
Initial Report FDA Received Date03/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-