• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IDEAL IMPLANT IDEAL IMPLANT STRUCTURED BREAST IMPLANT; SALINE-FILLED BREAST IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IDEAL IMPLANT IDEAL IMPLANT STRUCTURED BREAST IMPLANT; SALINE-FILLED BREAST IMPLANT Back to Search Results
Model Number 63501
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/31/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Deflation resulting in explantation.
 
Additional Manufacturer Narrative
The following updates were made to the report: date of this report, contact office - name/address/phone number, type of report, if follow-up, what type, unchecked evaluation summary attached, event problem and evaluation codes, and additional narratives/data.Evaluation summary was removed as an attachment.The evaluation summary is as follows: no manufacturing defect was found on macroscopic or microscopic examination of the explanted device.Debris/tissue was observed in the valve channel, which likely disrupted the seal, causing a leak and deflation.
 
Event or Problem Description
Deflation resulting in explantation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IDEAL IMPLANT STRUCTURED BREAST IMPLANT
Common Device Name
SALINE-FILLED BREAST IMPLANT
Manufacturer (Section D)
IDEAL IMPLANT
14881 quorum drive
suite 925
dallas TX 75254
MDR Report Key8450243
Report Number3011491947-2019-00080
Device Sequence Number1147734
Product Code FWM
UDI-Device Identifier10851795006121
UDI-Public(01)10851795006121(241)63501(17)200715
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P120011
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/15/2020
Device Model Number63501
Device Catalogue Number635
Device Lot Number62185
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2019
Initial Date Received by Manufacturer 01/03/2019
Supplement Date Received by Manufacturer01/03/2019
Initial Report FDA Received Date03/25/2019
Supplement Report FDA Received Date05/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age49 YR
-
-