OLYMPUS MEDICAL SYSTEMS CORP. VISERA CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-V2
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/19/2019
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to omsc but was returned to olympus (b)(4).(b)(4) sent the device to a third party laboratory for an additional microbiological testing.As a result of the testing, no microbe was detected from the sample collected from all channels of the device.The testing result cleared the (b)(6) guideline.Omsc reviewed the manufacturing history of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that as a result of microbiological testing by the user facility, the sample collected from the subject device tested positive for unspecified microbes ( > 100 cfu / 100ml).The device had been manually reprocessed using peracetic acid.There was no report of infection associated with this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA CYSTO-NEPHRO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key8450393
MDR Text Key139809489
Report Number8010047-2019-01358
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-V2
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/07/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/03/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-