• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. SILK SUTURE 30"(75CM) 0 BLK; SUTURE, NONABSORBABLE, SILK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. SILK SUTURE 30"(75CM) 0 BLK; SUTURE, NONABSORBABLE, SILK Back to Search Results
Catalog Number 680H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).To date, the device has not been returned.If the product is returned for evaluation, any further information derived from the evaluation will be submitted in a supplemental 3500a form.Additional information was requested and the following was obtained: was surgery delayed due to the reported event? no, action taken when event occurred? use another suture, was procedure successfully completed? yes.Were fragments generated? unknown, if yes, were they removed easily without additional intervention? unknown, patient status/ outcome / consequences - no, was other medical intervention - unknown.
 
Event or Problem Description
It was reported that the patient underwent a vats procedure on (b)(6) 2019 and suture was used.While the doctor was knotting the suture, it was broken.Another suture was used to complete the procedure.No adverse patient outcome was reported.
 
Additional Manufacturer Narrative
(b)(4).A manufacturing record review was performed for the finished device batch mjb935, 680h and no non-conformances were identified.Additional: an empty opened overwrap, an empty labeled winding former and four unopened samples of product code 680, lot mjb935 were returned for analysis.During the visual inspection of four samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, damaged or suture breakage were noted.A functional test was performed and the tensile strength force was above the minimum requirement.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of the representative samples, no breakage suture was found and the tested samples met the finished goods requirements.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SILK SUTURE 30"(75CM) 0 BLK
Common Device Name
SUTURE, NONABSORBABLE, SILK
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
MDR Report Key8450528
Report Number2210968-2019-79652
Device Sequence Number1341416
Product Code GAP
UDI-Device Identifier10705031008106
UDI-Public10705031008106
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date07/31/2023
Device Catalogue Number680H
Device Lot NumberMJB935
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2019
Initial Date Received by Manufacturer 02/26/2019
Supplement Date Received by Manufacturer04/26/2019
Initial Report FDA Received Date03/25/2019
Supplement Report FDA Received Date05/20/2019
Patient Sequence Number1
-
-