| Model Number |
SIG30CTAV |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during thoracoscopic lobectomy, at the first firing, the device was able to be used with no problem with jaws' opening and closing, approaching tissue, and until the closing of the jaws, but after pressing the green button, the handle could not be squeezed.A new cartridge was opened and used with the same handle, and then the device was able to be used with no problem.The surgical time was extended by less than 30 minutes.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the returned product noted that the reload was pre-fired and engaged in interlock.There were staples protruding from proximal staple cartridge channel.Functionally the reload was loaded into a post market vigilance instrument, the interlock was overridden, and the reload was applied to test media with proper staple placement and media transection.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the pre-fire condition with interlock engagement may occur if the instrument firing handle had been partially compressed and released after pressing the green firing button.In this situation, the safety interlock feature will engage and prevent the reload from firing a second time by ceasing the placement of staples and tissue transection and prevent patient harm.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|