• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® TWO PIN PLUG, MONITOR CABLE; TEMPERATURE SENSING CATHETER CABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® TWO PIN PLUG, MONITOR CABLE; TEMPERATURE SENSING CATHETER CABLE Back to Search Results
Model Number 153625H
Medical Device Problem Code Erratic or Intermittent Display (1182)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); Alteration In Body Temperature (2682)
Date of Event 03/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the body temperature was not indicated correctly.
 
Additional Manufacturer Narrative
The reported event was confirmed.A root cause could not be determined.The inspector received one temperature sensing cable with no packaging.It was noted that the connections (male connection) were bent.No missing, damaged or incorrect components are considered acceptable.The lot number is unknown; therefore, the device history record could not be reviewed.According to sec.801.116, medical devices having commonly known directions shall be exempt from section 502(f)(1) of the act insofar as adequate directions for common uses and are known to the ordinary individual.Product catalog number 153625h was deemed by appropriate subject matter experts (sme) to be within this definition.Section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the body temperature was not indicated correctly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® TWO PIN PLUG, MONITOR CABLE
Common Device Name
TEMPERATURE SENSING CATHETER CABLE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8452689
Report Number1018233-2019-01543
Device Sequence Number354786
Product Code EZL
UDI-Device Identifier00801741045905
UDI-Public(01)00801741045905
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K070582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,other
Initial Reporter Occupation Nurse Practitioner
Type of Report Initial,Followup
Report Date (Section B) 06/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number153625H
Device Catalogue Number153625H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2019
Initial Date Received by Manufacturer 03/04/2019
Supplement Date Received by Manufacturer05/30/2019
Initial Report FDA Received Date03/26/2019
Supplement Report FDA Received Date06/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-