|
Results: the pet lock was damaged but intact at the proximal end of the pusher assembly.The pusher assembly was fractured approximately 34.0 cm from the proximal end.The pusher assembly was kinked approximately 9.0 cm, 32.5 cm, 35.5 cm, 70.0 cm, 87.0 cm, 111.0 cm and 141.5 cm from the proximal end.The embolization coil was detached from the pusher assembly distal detachment tip (ddt).The embolization coil was undamaged.The introducer sheath was undamaged.The pull wire within the pusher assembly was fractured approximately 170.5 cm from the proximal end.Conclusions: evaluation of the returned pod14 confirmed a fractured pusher assembly and detached embolization coil.It was reported that the physician felt resistance while advancing the pod14 into the lantern.If the device is forcefully advanced against resistance, damage such as a kink may occur.Subsequently, if the kinked device is further manipulated, damage such as a fracture may occur.If the pusher assembly becomes fractured and two segments are separated, the pull wire may be retracted out of the distal detachment tip (ddt) and the embolization coil will detach from its pusher assembly.Based on the reported complaint, the root cause of the initial resistance experienced during the procedure could not be determined.Further evaluation revealed kinks along the pusher assembly and a fractured pull wire.Some of these damages were likely incidental to the reported complaint and may have occurred during packaging for return to penumbra.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|