| Catalog Number |
2296003225 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/27/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Awaiting device evaluation.A follow up report will be filed once the quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that two blades broke at the mount during a high tibial osteotomy.It was also reported that there was a delay of a few minutes to obtain a replacement blade.It was further reported that the procedure was completed successfully and there were no adverse consequences as a result of this event.This report is for the first blade that broke.
|
| |
|
Event or Problem Description
|
|
It was reported that two blades broke at the mount during a high tibial osteotomy.It was also reported that there was a delay of a few minutes to obtain a replacement blade.It was further reported that the procedure was completed successfully and there were no adverse consequences as a result of this event.This report is for the first blade that broke.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results indicate that the blade mount on both sides were possibly exposed to a torsional loading while cutting.This type of loading could occur when applying a bending moment to the blade while in use.The ifu of handpieces associated with this device warn the user against this type of use: "do not apply excessive pressure, such as bending or prying, with the blade.Excessive pressure may bend or fracture the blade and cause tissue damage and/or loss of tactile control".The quality investigation is complete.
|
| |
|
Search Alerts/Recalls
|