• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO PRECISION THIN (9.0 X 0.38 X 31.0MM); BLADE, SAW, GENERAL & PLASTIC SURGERY,SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO PRECISION THIN (9.0 X 0.38 X 31.0MM); BLADE, SAW, GENERAL & PLASTIC SURGERY,SURGICAL Back to Search Results
Catalog Number 2296003225
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/27/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Awaiting device evaluation.A follow up report will be filed once the quality investigation is complete.
 
Event or Problem Description
It was reported that two blades broke at the mount during a high tibial osteotomy.It was also reported that there was a delay of a few minutes to obtain a replacement blade.It was further reported that the procedure was completed successfully and there were no adverse consequences as a result of this event.This report is for the second blade that broke.
 
Additional Manufacturer Narrative
Investigation results indicate that the blade mount on both sides were possibly exposed to a torsional loading while cutting.This type of loading could occur when applying a bending moment to the blade while in use.The ifu of handpieces associated with this device warn the user against this type of use: "do not apply excessive pressure, such as bending or prying, with the blade.Excessive pressure may bend or fracture the blade and cause tissue damage and/or loss of tactile control".The quality investigation is complete.
 
Event or Problem Description
It was reported that two blades broke at the mount during a high tibial osteotomy.It was also reported that there was a delay of a few minutes to obtain a replacement blade.It was further reported that the procedure was completed successfully and there were no adverse consequences as a result of this event.This report is for the second blade that broke.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION THIN (9.0 X 0.38 X 31.0MM)
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY,SURGICAL
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key8453345
Report Number0001811755-2019-00946
Device Sequence Number193622
Product Code GFA
UDI-Device Identifier04546540046550
UDI-Public04546540046550
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/01/2023
Device Catalogue Number2296003225
Device Lot Number18347027
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2019
Initial Date Received by Manufacturer 02/27/2019
Supplement Date Received by Manufacturer05/09/2019
Initial Report FDA Received Date03/26/2019
Supplement Report FDA Received Date05/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-