| Catalog Number |
75018942 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
03/05/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this is 3rd revision of hip (previous product experience).Patient dislocated hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Results of investigation: a dislocation/ luxation was reported with a polarcup xlpe insert 22/43.The device in question was not sent back for investigation and a product evaluation could not be conducted.A review of the batch record revealed no deviations from the standard manufacturing process that could have contributed to the reported failure mode.A review of the complaint history revealed no other complaints for the batch in question.One other complaint with similar issue has been reported for the device.No supporting clinical documents were provided to conduct a thorough medical investigation.Therefore based on insufficient information, a thorough clinical assessment could not be performed.The stated failure mode could not be confirmed.The root cause of reported dislocation/ luxation cannot clearly be reproduced and stays undetermined.Based on available information the need for corrective action is not indicated.The complaint will be reopened should additional information be received.Smith and nephew will continue to monitor this device for similar issues.
|
| |
|
Event or Problem Description
|
|
It was reported that this is the third revision due to hip dislocation.
|
| |
|