• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESHAPE LIFESCIENCES LAP-BAND AP STANDARD W/ ACCESS PORT I; ADJUSTABLE GASTRIC BAND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESHAPE LIFESCIENCES LAP-BAND AP STANDARD W/ ACCESS PORT I; ADJUSTABLE GASTRIC BAND Back to Search Results
Model Number B-2240
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/25/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Taper ii.Device evaluation summary: the device was returned to apollo for analysis, and visual examination confirmed the connector type as taper ii.The band tubing was noted to be separated near the port tubing ss connector.Needle marks were noted on the port septum and scratches were observed on the port housing and port base.A port leak test was performed and leakage was observed from an opening on the port tubing at the taper tubing junction.An air leak test was not feasible as only the access port was returned.A fill inspection test was performed and no blockage was noted.Under microscopic analysis, the opening on the port tubing, was noted to have a smooth edge, consistent with wear and thinning.The end of the band tubing, separated approximately.6 inches past the ss connector, was noted to have striated edges consistent with damage from a surgical end cut to remove the device.Device labeling addresses the reported event as follows: precautions: care must be taken during band adjustment to avoid puncturing the tubing that connects the access port and band, as this will cause leakage and deflation of the inflatable section.Failure to create a stable, smooth path for the access port tubing, without sharp turns or bends, can result in tubing breaks and leakage.In order to avoid incorrect placement, the port should be placed lateral to the trocar opening.A pocket must be created for the port so that it is placed far enough from the trocar path to avoid abrupt kinking of the tubing.The tubing path should point in the direction of the access port connector so that the tubing will form a straight line with a gentle arching transition into the abdomen.Adverse events: it is important to discuss all possible complications and adverse events with your patient.Complications which may result from the use of this product include the risks associated with the medications and methods utilized in the surgical procedure, the risks associated with any surgical procedure and the patient's degree of intolerance to any foreign object implanted in the body.Unplanned deflation of the band may occur due to leakage from the band, the port or the connecting tubing.Warnings: patients should be advised that the lap-band system is a long-term implant.Explant (removal) and replacement surgery may be indicated at any time.Medical management of adverse reactions may include explantation.Revision surgery for explantation and replacement may also be indicated to achieve patient satisfaction.
 
Event or Problem Description
Reported as: a patient with the lap-band system was reported to have "lap-band port malfunctioned and was removed.Port had fractured and was broken were it connects to tubing.Was replaced with a new port".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAP-BAND AP STANDARD W/ ACCESS PORT I
Common Device Name
ADJUSTABLE GASTRIC BAND
Manufacturer (Section D)
RESHAPE LIFESCIENCES
1001 calle amanecer
san clemente CA 92673
MDR Report Key8453990
Report Number3013508647-2019-00010
Device Sequence Number10811210
Product Code LTI
UDI-Device Identifier10811955020183
UDI-Public10811955020183
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P000008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/24/2011
Device Model NumberB-2240
Device Catalogue NumberB-2240
Device Lot Number1787725
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2019
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/25/2019
Initial Report FDA Received Date03/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age44 YR
Patient Weight77
-
-