Model Number MMT-523NAS
Medical Device Problem Codes Mechanical Problem (1384); No Apparent Adverse Event (3189)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 03/04/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that they experienced high blood glucose.The customer¿s blood glucose level was over 600 mg/dl.The customer was treated with insulin.The customer declined troubleshoot for high blood glucose.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device passed the rewind, basic occlusion, prime and displacement test.Device received with stuck motor error alarm loop during bolus delivery.The motor was tested outside of the device and passed.The motor may have had and intermittent failure that was not detected during our testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAS
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8454248
Report Number3004209178-2019-72227
Device Sequence Number10812633
Product Code OYC
UDI-Device Identifier00613994743886
UDI-Public(01)00613994743886
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-523NAS
Device Catalogue NumberMMT-523NAS
Device Lot NumberA3523NASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2019
Initial Date Received by Manufacturer 03/18/2019
Supplement Date Received by Manufacturer04/16/2019
Initial Report FDA Received Date03/26/2019
Supplement Report FDA Received Date04/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age60 YR
Patient Weight66
-
-