| Model Number |
MMT-523NAS |
| Medical Device Problem Codes |
Mechanical Problem (1384); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
03/04/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that they experienced high blood glucose.The customer¿s blood glucose level was over 600 mg/dl.The customer was treated with insulin.The customer declined troubleshoot for high blood glucose.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the rewind, basic occlusion, prime and displacement test.Device received with stuck motor error alarm loop during bolus delivery.The motor was tested outside of the device and passed.The motor may have had and intermittent failure that was not detected during our testing.
|
| |
|