Model Number EX1150
Device Problem Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/07/2019
Event Type  malfunction  
Manufacturer Narrative
The field complaint of the transmitter's power light not working and a burn smell coming from the contacts was verified. Final analysis found burnt components on the main printed circuit board (pcb) and on its¿ inner casing. Further testing revealed that the burnt damage to the original ac power adapter was the cause of the no power issue. It was concluded that the transmitter was hit by a high current flow through the ac wall power outlet into the ac power adapter, thus damaging its¿ main pcb.
 
Event Description
It was reported that the merlin@home transmitter¿s power light was not working. The transmitter would not power on. Transmitter's prongs and contacts were tested. A burn smell was coming from the electrical adapter/ contacts. Transmitter was returned and replaced. There were no adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMERLIN@HOME, RF TELEMETRY BASIC
Type of DevicePACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8454308
MDR Text Key139956540
Report Number2938836-2019-02166
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial
Report Date 03/26/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberEX1150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2019
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received03/11/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-