MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAP; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751NAP
Medical Device Problem Code Device Difficult to Program or Calibrate (1496)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/29/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Insulin pump received with no button response due to flattened dome switch on esc, act, up arrow and down arrow button.The connector was inspected and locked on lcd board.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
 
Event or Problem Description
The customer reported via phone call that the insulin pump buttons were sticking.The customer¿s blood glucose level was 195 mg/dl.Customer was advised to observe time clock for one minute to verify if time advances and they stated that the time was advances.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751NAP
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8455166
Report Number3004209178-2019-72758
Device Sequence Number354803
Product Code OZO
UDI-Device Identifier00643169736092
UDI-Public(01)00643169736092
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-751NAP
Device Catalogue NumberMMT-751NAP
Device Lot NumberA6751NAPJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2019
Initial Report FDA Received Date03/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age72 YR
Patient Weight62
-
-