| Model Number |
MMT-551LNAP |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Insulin pump received with no button response due to flattened esc and act button dome switch.The keypad connector on lcd is locked properly during visual inspection.Unable to perform self test and displacement test due to no button response.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the insulin pump had a keypad anomaly.The customer¿s blood glucose was unknown at the time of incident.Customer stated that the insulin pump buttons were hard to press.No significant events leading to keypad anomaly were observed.Keypad anomaly troubleshooting was performed and confirmed that time is advancing.The customer was advised to discontinue use of pump and revert to back-up plan.The insulin pump is not expected to return for analysis.
|
| |
|