| Catalog Number |
234601000 |
| Medical Device Problem Codes |
Material Deformation (2976); Material Twisted/Bent (2981); Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
| Date of Event |
03/06/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device is an instrument and is not implanted/explanted.
|
| |
|
Event or Problem Description
|
|
It was reported that the screw driver was damaged and stripped.There was 5 minutes surgical delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: the instrument associated with this report was not returned.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # = > (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthese is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.Corrected: h6.Product complaint # : (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.The root cause is attributed to wear out.The investigation could not verify or identify any product contribution to the reported event with the information provided.Based on the investigation, the need for corrective action is not indicated.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|