• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MITG-OKLAHOMA CITY TRUCLEAR; HYSTEROSCOPE (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MITG-OKLAHOMA CITY TRUCLEAR; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Model Number 7209807S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation summary: one device was received for evaluation.The returned sample did not meet specification as received.The reported condition was confirmed.The most likely cause was determined to be related to the user misuse.The unit serial number is below rohs cutoff, and will be scrapped.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during reprocessing, device had broken connector.No patient involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUCLEAR
Common Device Name
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
MITG-OKLAHOMA CITY
75 s. meridian ave
oklahoma city OK 73107
MDR Report Key8457193
Report Number1643264-2019-20003
Device Sequence Number345071
Product Code HIH
UDI-Device Identifier10884521749900
UDI-Public10884521749900
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K132015
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 03/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2019
Device Model Number7209807S
Device Catalogue Number7209807S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2019
Initial Report FDA Received Date03/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-